If the MiSeq fails to measure flow rate when starting a run, perform the following steps to resolve the issue. Steps 1 and 2 can be performed together, before repeating the flow check.
- Inspect the flow cell and gasket for any signs of damage:
- Open the MiSeq door and remove the flow cell.
- Confirm that the flow cell gasket is flush as shown in the following images.

- If the gasket is raised or distorted, do not use this flow cell. Email Technical Support to recieve a replacement for the defective flow cell. Technical Support will ask for the lot and serial numbers of the affected flow cell, and may ask for a picture of the defective gasket for quality control. Proceed with the pre-run checks using a new flow cell from another kit, if available.
- If the gasket is flush with the plastic cartridge, clean, and reseat the flow cell.
- Clean and reseat the flow cell:
- Lightly rinse the flow cell with laboratory-grade water until both the glass and plastic cartridge are thoroughly rinsed of excess salts.
- Using care around the black flow cell port gasket, thoroughly dry the flow cell and cartridge with a lint-free lens cleaning tissue. Gently pat dry in the area of the gasket and adjacent glass.
- Clean the flow cell glass with an alcohol wipe, but do not use alcohol on the gasket.
- Check the flow cell stage for any debris or salt and clean the flow cell stage using an alcohol wipe or a lint-free tissue moistened with ethanol.
- Return the flow cell to the stage, and make sure that it is seated properly.
- Using MiSeq Control Software, perform another flow check.
- If the flow check still fails, power cycle the instrument:
- Close MiSeq Control Software, shut down Windows, and turn off the instrument using the power switch on the back panel.
- Leave the instrument off for at least 2 minutes, to allow all parts to completely power down.
- Turn on the instrument using the power switch on the back panel.
- Using MiSeq Control Software, perform another flow check.
- If the flow check still fails, use a different flow cell (eg, from another kit):
- Open the MiSeq door and remove the flow cell.
- Clean the flow cell as described in step 2.
- Place the new flow cell on the stage, and make sure that it is seated properly.
- Using MiSeq Control Software, go back in the run set up to repeat the flow cell RFID check, this will record the new consumable information.
- Continue with the run set up and perform another flow check.
- If the flow check passes, record the lot number of the flow cell and email Technical Support to receive a replacement for the defective flow cell.
- If the issue persists with a new flow cell, email Technical Support for further assistance.