How many cycles of SBS chemistry are in my kit?

01/25/2022


Making sure that sufficient volumes of SBS reagents are loaded onto the sequencing instrument is crucial for a successful run. Running out of reagents during a run impacts intensity and quality, and can result in a run failure. Dual-index sequencing, depending on the instrument, follows different workflows, as shown in the Indexed Sequencing Overview Guide. On certain instruments, dual-index runs need an additional seven cycles of reagents for the chemistry-only cycles at the beginning of the second Index Read. The seven chemistry-only cycles are required for the workflows in which the second Index Read sequencing occurs after the forward template anneals to the grafted P5 primer on the surface of the flow cell. These seven cycles must be considered when calculating the number of cycles supported by the kit.

The following table outlines available reagent kit sizes and the maximum number of cycles each kit can perform, taking the additional seven cycles into consideration:

Reagent Type Reagent Kit Size Maximum Number of Cycles Additional Cycles Needed for Dual Index?
iSeq 100 i1 (v1 or v2) 300 322 No
MiniSeq Rapid Kit 100 128 Yes - 7 cycles
High Output or
Mid Output
75 92 No
150 168
300 318
MiSeq v2
(including Micro and Nano kits)
50 79 Yes - 7 cycles
300 329
500 529
v3 150 179
600 629
NextSeq 500/550 High Output or Mid Output 75 92 No
150 168
300 318
HiSeq 1000/1500/2000/2500 Rapid SBS v2 50 79 7 cycles required for paired-end flow cells
200 229
500 529
TruSeq SBS v3 50 58
200 209
HiSeq SBS v4 50 79
250 279
NextSeq 1000/2000
P1 100 138 No
600 638
P2 100 138 No
200 238
300 338*
600 638
P3 50 88 No
100 138
200 238
300 327*
HiSeq 3000/4000 N/A 50 79 7 cycles required for single-read flow cells
150 179
300 329
NovaSeq 6000 SP v1.0 100 130 Yes - 7 cycles
200 230
300 330
500 530
S1 and S2 v1.0 100 130
200 230
300 330
S4 v1.0 200 230
300 330
SP v1.5 100 138 No
200 238
300 338
500 538
S1 and S2 v1.5 100 138
200 238
300 338
S4 v1.5 35 72
200 238
300 338

 

*The NextSeq 1000/2000 P1 and P2 kit has 338 total cycles while the P3 v3 kit has 327 total cycles.

For more information about each SBS reagent kit, including how to prepare, combine, and split SBS reagents when possible, visit the support page for each kit; also refer to the bulletin Maximum read length for Illumina sequencing platforms for maintaining high data quality.