An issue has been identified that impacts NextSeq Control Software v4.0.1 run set up with Local Run Manager on NextSeq 500/550 instruments. This causes the run to stop several seconds after starting the run with an “Invalid Role” error.


The error occurs when User Management is enabled in Local Run Manager and the user initiating the run in NextSeq Control Software has a "User" level role, instead of an "Admin" level role. Determine role level by navigating to Local Run Manager Dashboard > Menu > Tools > User Management.

This issue affects only systems running NextSeq Control Software v4.0.1 on NextSeq instruments or the NextSeq 550Dx instruments run in research mode.
Illumina developers are aware of the issue and are working on a solution.
Currently, there are two workarounds:
Since the error occurs only a few seconds after starting the run, the loaded reagents are not marked as used and can be used for the new run set up.