The options are based on the available DRAGEN pipelines:
Analysis is available through the on-board DRAGEN Bio-IT Platform or on the cloud using BaseSpace Sequence Hub. The following DRAGEN pipelines are available:
For each read, entering a minimum of 26 cycles and a maximum of 151 cycles helps ensure data quality. The exact number of cycles depends on your experiment. NextSeq 1000/2000 Control Software requires at least 1 cycle for Read 1, but displays a warning when the number of cycles in Read 1 is less than 26. The total number of cycles for Read 1, Index 1, Index 2, and Read 2 cannot be greater than the number of cycles supported by the kit plus 27. NextSeq 1000/2000 Control Software will display a warning when Index 1 and Index 2 are less than 6 cycles. The warning will not be shown if Index 1 or Index 2 is 0 cycles.
Run duration depends on the type of reagents and number of cycles. For example run times, see Specifications for the NextSeq 1000 and NextSeq 2000 Systems.
The NextSeq 1000/2000 P2 Reagents can generate up to 120 Gb per sequencing run and the NextSeq 2000 P3 Reagents can generate up to 330 Gb per sequencing run. For detailed performance parameters, see Specifications for the NextSeq 1000 and NextSeq 2000 Systems.
Depending on the sequencing mode selected, sample information is entered using Instrument Run Setup BaseSpace Sequence Hub or a Sample Sheet v2 template (see Product Files).
The following applications are targeted for the NextSeq 1000/2000 system:
The loading volume is 20 µl and the loading concentration varies by library type:
Library Type | Loading Concentration (pM) |
AmpliSeq™ for Illumina Library PLUS | 750 |
Illumina DNA Prep | 750 |
Illumina DNA Prep with Enrichment | 1000 |
Illumina Stranded Total RNA with Ribo-Zero Plus | 750 |
Illumina Stranded mRNA Prep | 750 |
Illumina DNA PCR-Free | 1000 |
100 % PhiX | 650 |
TruSeq DNA Nano 350 | 1200 |
TruSeq DNA Nano 550 | 1500 |
TruSeq Stranded mRNA | 1000 |
For other library types, 650 pM is the recommended starting loading concentration. Optimize this concentration over subsequent runs to identify a loading concentration that consistently yields data that meets specifications.
The number of libraries you can sequence at one time depends on the desired coverage. Using supported Illumina kits, you can multiplex up to 384 libraries. For coverage considerations, see the Sequencing Coverage Calculator.
Custom primers are supported for the NextSeq 1000/2000 P2 Reagents kit and the NextSeq 1000/2000 P3 Reagents kit. If using the NextSeq 1000/2000 P2 (v2) Reagents kit (catalog # 20043738, 20043737, and 20043736), you will need to use a custom recipe to use custom primers. See Product Files for more information.
Yes, NextSeq 1000/2000 P2 Reagents and NextSeq 1000/2000 P3 Reagents include all consumables to support both single-read and paired-end sequencing. During run setup, you can configure the run to generate either single-read or paired-end data.